The ORCA-Flash4.0 V2 offers unrivaled flexibility across a wide range of imaging applications. Easily change from USB 3.0 to Camera Link connectivity. Switch from a blazing fast scan to a virtually noiseless slow scan by a simple click in software. Use our Lightsheet Readout Mode™ for seamless integration with light sheet microscopy systems. Robust triggering allows the ORCA-Flash4.0 V2 to drive other devices or be driven by them. And then there’s the highest quantum efficiency (QE) of any sCMOS camera on the market.
- Standard scan mode with 1.9 electrons rms (1.3 electrons median) read noise
- Slow scan mode with 1.5 electrons rms (0.9 electrons median) read noise
- Lightsheet Readout Mode™ for light sheet microscopy
- 4.0 megapixels with 6.5 x 6.5 µm pixel size
- Frame rates
- 100 frames/s at full resolution with Camera Link
- 30 frames/s at full resolution with USB 3.0
- Over 70% QE at 600 nm
The ORCA-Flash4.0 V2 features two scan modes: a blazing fast scan (standard) or a virtually noiseless slow scan. The fast standard scan produces a read noise of 1.9 electrons rms (1.3 electrons median), which is suitable for most experiments. The slow scan with read noise of 1.5 electrons rms (0.9 electrons median) is designed for experiments where lower noise is more important than raw speed. Both the USB 3.0 and Camera Link configurations of the camera have this low noise capability.

To enable the best speeds and synchronization for light sheet microscopy, the ORCA-Flash4.0 V2 configured with the Camera Link interface can be read out using the Lightsheet Readout Mode™ feature. In this mode, the camera’s Gen II sCMOS sensor is read out in one sweep across the sensor from top to bottom or bottom to top.
Every ORCA-Flash4.0 V2 includes ImageConductor Connectivity™, so that it’s enabled for both USB 3.0 (default) and high-speed Camera Link. If your imaging tempo is 30 frames/s, then the default configuration with USB 3.0 is right for you. If you need faster speed, just add a Camera Link board now or later to achieve 100 frames/s at full resolution. Both options deliver the same low noise, high QE imaging for unprecedented sensitivity.
| Readout Speed |
Camera Link |
USB 3.0 |
| Horizontal Pixel |
Binning |
Horizontal Pixel |
Binning |
| 2048/1536/1024/512 |
2x2, 4x4 |
2048/1536/1024 |
512 |
2x2, 4x4 |
| Vertical Line |
2048 |
100 |
30 |
100 |
100 |
| 1024 |
200 |
60 |
200 |
200 |
| 512 |
400 |
120 |
400 |
400 |
| 256 |
801 |
240 |
801 |
801 |
| 128 |
1603 |
481 |
1603 |
1603 |
| 64 |
3206 |
968 |
3206 |
3206 |
| 8 |
25,655 |
7894 |
25,655 |
25,655 |
| Readout speed at center position (frames/s, typical) |
The ORCA-Flash4.0 V2 has an added Global Reset function, so you can acquire global exposures and choose to have either an external source or the camera be master of the timing.
Thanks to its low noise and high QE values across a wide range of wavelengths, the ORCA-Flash4.0 V2 has the sensitivity required for many fluorescence microscopy techniques. These imaging techniques include super-resolution microscopy, ratio imaging, FRET, high-speed calcium ion imaging, TIRF microscopy, real-time confocal microscopy, and light sheet microscopy. |
 |
With 4.0 megapixels at 6.5 µm x 6.5 µm each, the ORCA-Flash4.0 V2 offers a field of view 2.5 times larger than a standard EM-CCD.
In addition, this camera’s small pixels can resolve much finer details than the larger pixels of an EM-CCD camera.
Sample:FluoCells Prepared Slide #1
Objective lens:S Plan Fluor 100x |
 |
Knowing as much as possible about your camera helps increase confidence in the results it produces, especially under demanding experimental conditions. Every ORCA-Flash4.0 V2 is individually characterized at the factory before it ships, and the results of these tests are included with each camera. A measured noise histogram, photon transfer curve, rms noise value and conversion factor (electron/count) are provided along with simple formulas to make use of this information.
- Super-resolution microscopy
- TIRF microscopy
- Ratio imaging
- FRET
- High-speed calcium ion imaging
- Real-time confocal microscopy
- Light sheet microscopy
- PC Requirements. With the introduction of the ORCA-Flash 4.0 V2, users are now able to stream
4 megapixel images to their computers 100 frames per second. The computer requirements for this
high data rate can be met by using the guidelines listed this PC Recommendations for ORCA-Flash4.0 V2.
- Water Chiller. The following water chiller(s) have been qualified by Hamamatsu for use with the
C11440-22C ORCA-Flash4.0 V2: Julabo F25-ED Refrigerated/Heating Circulator.
If you have any questions please contact our Technology Group at 908-231-0960 or via the web. |